TOY18FIX Dual UV Detector
2 channel
200-600
FIX
Dual UV detector version measures at two fixed wavelengths which are adjusted in production according to customer order in the ranges of 200 – 600 nm.
Related variants
Categories
Manufacturer
ECOM spol. s r.o.
Part number
TOY2200Y
- Description
- Parameters
- Files to download 1
The unit is supplied with an AMY17 PCB board. It powers the lamps and controls the unit. It has a lot of new functions including higher-speed sampling.
Most important features of TOY18FIX detectors:
•Compact dimensions
•Built-in lamp working hours counter
•Easy replacement of flow cell
•Sophisticated diagnostic software
Number of channels (Signals) | Two fix wavelengths in range of 200 – 600nm |
---|---|
Number of CCD elements | 256 |
Communication | RS232 |
Accuracy of adjustment | ± 1 nm |
Typical spectral half-width | 8 nm |
Reproducibility | ±0.5 nm |
Light source | Deuterium lamp |
Noise (Test cell, 254nm, TC 2s, 10 Hz) | ± 5 x 10-5 AU |
Drift (Test cell, 254nm) | 1 x 10-3 AU/h |
Time constant | 0.5 s, 0.75 s, 1.0 s, 2.0 s, 4.0 s, 8.0 s, 16.0 s, 0.2 s, 0.1 s |
Wetted materials | FEP; fused silica, stainless steel |
Maximal flow rate | 500 or 3 000 ml/min depending on the cell |
Communication | RS232; optionally USB and LAN |
Power supply | 24 -36 V DC |
Power input | 48 W |
Optical cables | no |
Dimensions W x H x D | 131 x 167 x 386 mm (5.15 x 6.57 x 15.20 in) |
Weight | 3.20 kg (7.06 lb) |