Flash 06S Dual Detector with housing
Two wavelengths simultaneously, according chosen interference filters in range 200 - 600 nm
- Description
- Parameters
Flash 06S DUAL is a dual detector with two interference filters. Detector can be adjusted using interference filters at two fixed wavelengths in range of 200 – 600 nm. Noise level at 254 nm is ± 2 x 10-5 AU with an empty cell. (Filters 254 nm and 280 nm include price, other filters on request) High standard deuterium lamp in a special socket enables easy changing.
This unit is used in liquid chromatography in flash and preparative applications. By using analytical cell can be detector used also for analytical purpose.
The unit is designed as stand alone with 90–264 V power supplyand RS232 controlling. It is derived from OEM build-in unit with powering 24-36 V DC.
The unit offers many advantages:
- measuring at two wavelengths simultaneously
- wavelength setting from 200 up to 600 nm according used interference filter
- lamp work hours are counted using the built-in counter
- the cell is easy to replace from the side of the detector
- powering by 90–264 V
- unit is controlled by RS232 interface
Preparative cell PLCC 15 (supplied with unit):
volume/optical path(adjustable) | 45 μl / 0.3 mm (basic) |
55 μl / 1.4 mm | |
70 μl / 2.4 mm | |
connecting | tubing with OD = 1/8", thread ¼“-28 |
maximal flow rate | 500 ml/min |
Typical spectral half-width | According used interference filter |
---|---|
Light source | Deuterium discharge lamp |
Output for integrator | 1 V/AU (in digital form only) |
Time constant (T90) | 0 s; 0.3 s; 0.6 s; 1.5 s |
Communication | RS232 |
Power supply | 90–264 V |
Power input | 90 VA |
Noise level at empty cell (254 nm, TC 1 s) | ± 2 x 10-5 AU |
Drift at empty cell (254 nm after 1 h) | 1 x 10-3 AU/hr |
Materials in contact with mobile phase | PTFE; fused silica, stainless steel |
Dimensions W x H x D | 220 x 170 x 450 mm |
Weight | 7.9 kg |